Comparison of Smartphone Ophthalmoscopy With Slit-Lamp Biomicroscopy for Grading Vertical Cup-to-Disc Ratio.

Russo Andrea, Mapham William, Turano Raffaele, Costagliola Ciro, Morescalchi Francesco, Scaroni Nicolò, Semeraro Francesco

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE OF THE STUDY: The purpose of the study was to determine the agreement between smartphone ophthalmoscopy and slit-lamp indirect biomicroscopy when assessing vertical cup-to-disc ratios (VCDRs).

MATERIALS AND

METHODS

This was a clinical-based, prospective, comparative instrument study performed in 110 patients with ocular hypertension (OH) or primary open angle glaucoma (POAG). Patients underwent estimation of VCDR by undilated smartphone ophthalmoscopy and slit-lamp biomicroscopy by 2 masked glaucoma specialists.

RESULTS

The differences between the mean VCDR estimations obtained by each techniques were not statistically significant. Overall exact agreement between the 2 modalities was found in 21 of 29 eyes (72.4%; simple κ=0.63, confidence interval, 0.52-0.73, P<0.001) in POAG patients and in 52 of 78 eyes (66.7%) in OH patients. The optic nerve head was not gradable with smartphone ophthalmoscopy in 1 eye with POAG and in 2 eyes with OH because of media opacities and/or small pupil diameter.

CONCLUSIONS

Smartphone ophthalmoscopy showed substantial agreement with slit-lamp examination for the estimation of the VCDR. The ubiquitous diffusion of the smartphones, together with their connectivity and portability features, enables an extensive benefit for this technology to be used in glaucoma screening, especially in low-resource settings.

Для цитирования: Russo Andrea, Mapham William, Turano Raffaele, Costagliola Ciro, Morescalchi Francesco, Scaroni Nicolò, Semeraro Francesco. Comparison of Smartphone Ophthalmoscopy With Slit-Lamp Biomicroscopy for Grading Vertical Cup-to-Disc Ratio.. Journal of glaucoma. 2016. DOI: 10.1097/ijg.0000000000000499.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.