Medical principles and practice : international journal of the Kuwait University, Health Science Centre · 2016

Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria.

Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Onochie Chizoba, Okoye Obiekwe, Ogunro Adunola, Aribaba Tade, Hassan Kunle, Onakoya Adeola

Аннотация

OBJECTIVE: The aim of this study was to compare intraocular pressure (IOP) measured with the Tono-Pen® to that measured using the Goldmann applanation tonometer (GAT) in patients with primary open angle glaucoma attending a glaucoma clinic.

SUBJECTS AND METHODS: A comparative clinic-based observational study was conducted involving 75 patients (39 men and 36 women) attending a glaucoma clinic in Southwest Nigeria. A pretested structured questionnaire was used to collect sociodemographic data. The Tono-Pen and the GAT were used to measure the IOP in each patient. Central corneal thickness (CCT) was measured with an ultrasonic pachymeter, and the corrected GAT value was calculated for each patient. The Tono-Pen readings were compared with the uncorrected and corrected GAT readings. Descriptive and comparative analyses were performed. Values for p < 0.05 were considered statistically significant.

RESULTS: The mean age of the participants was 60.39 ± 16.71 years. The mean IOP using the Tono-Pen was 21.1 ± 6.8 mm Hg versus the mean uncorrected GAT value of 17.1 ± 6.9 mm Hg and the corrected GAT value of 18.9 ± 7.5 mm Hg. The mean CCT was 510.5 ± 29.6 µm. The mean differences between the Tono-Pen reading and uncorrected and corrected GAT readings were 3.9 ± 2.6 and 2.1 ± 3.5 mm Hg, respectively. Gender (Tono-Pen vs. GAT; p = 0.981 vs. 0.437) and corneal thickness (p = 0.057) did not significantly affect the IOP value. Of the 75 patients, 68 (90.7%) preferred the Tono-Pen to the GAT.

CONCLUSION: In this study, the Tono-Pen gave a higher value for IOP than the uncorrected and corrected GAT values. Gender and corneal thickness did not significantly affect the measurements. Most patients found the Tono-Pen more acceptable than the GAT.