MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/j.jcrs.0000000000001536

Медицинская библиотека · Journal of cataract and refractive surgery · 2024

Characterization of vault with two models of posterior chamber phakic intraocular lenses.

Paredes Beatriz, Mora de Oñate Javier, Martín Sánchez Dolores, Piñero David P

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate and compare dynamic vault range (DVR) and asymmetry of the vault during a 3-month follow-up after the implantation of 2 posterior chamber phakic intraocular lenses (pIOLs).

SETTING

Aver Clinic, Madrid, Spain.

DESIGN

Prospective comparative study.

METHODS

119 eyes (65 patients) that underwent refractive surgical correction with implantation of 1 of 2 distinct pIOLs were enrolled: Eyecryl Phakic from Biotech Vision Care (Eyecryl group, 72 eyes) and Evo Visian Implantable Contact Lens from Staar Surgical (ICL group, 47 eyes). Besides evaluation of visual acuity, refraction, and ocular integrity, the pIOL vault centrally and at 2 mm nasally and temporally as well as the DVR from photopic (50 lux) to mesopic (10 lux) illumination conditions were measured.

RESULTS

No significant differences were found between the pIOL groups in visual and refractive outcomes ( P ≥ .454). No significant differences between the groups were found in central (523.72 ± 168.4 vs 494.16 ± 156.7 μm, P = .248) and temporal (499.43 ± 155.8 vs 431.28 ± 150.5 μm, P = .067) vaults. However, nasal vault was significantly lower in the ICL group (465.6 ± 149.1 vs 375.4 ± 144.0 μm, P = .045). A trend of a larger DVR was observed in the ICL group, although differences between groups did not reach statistical significance (54.00 ± 36.39 vs 86.5 ± 57.9 μm, P = .070). The pIOL diameter only correlated significantly with vault measurements in the ICL group ( r ≥ 0.650, P < .001).

CONCLUSIONS

The Eyecryl pIOL showed more symmetric vaults and a trend of fewer light-induced changes in the central vault compared with the ICL pIOL. The clinical relevance of this finding should be investigated further.

Для цитирования: Paredes Beatriz, Mora de Oñate Javier, Martín Sánchez Dolores, Piñero David P. Characterization of vault with two models of posterior chamber phakic intraocular lenses.. Journal of cataract and refractive surgery. 2024. DOI: 10.1097/j.jcrs.0000000000001536.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.