Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2007

Reduction of ventricular tachyarrhythmia by treatment of atrial fibrillation in ICD patients with dual-chamber implantable cardioverter/defibrillators capable of atrial therapy delivery: the REVERT-AF Study.

Gradaus Rainer, Seidl Karlheinz, Korte Thomas, Himmrich Ewald, Wieneke Heinrich, Schuchert Andreas, Bauer Wolfgang, Gerss Joachim, Wollmann Christian G, Borggrefe Martin, Böcker Dirk

СтатьяМногоцентровое исследованиеРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

AIMS: The purpose of this prospective, randomized, multicentre study was to investigate whether the incidence of ventricular tachyarrhythmia can be reduced in standard implantable cardioverter/defibrillator (ICD) patients by implanting a dual-chamber ICD capable of atrial therapy delivery.

METHODS AND

RESULTS

A Jewel AF or GEM III AT ICD (Medtronic Inc., Minneapolis, MN, USA) was implanted in 122 patients (62.3 +/- 10.5 years; 84.4% male; coronary artery disease 71.3%; left ventricular ejection fraction 39.7 +/- 13.6%; secondary ICD indication 91%). Overall, 31.2% of the patients had paroxysmal atrial fibrillation (AF)/atrial tachycardia (AT) before ICD implantation (n = 38). Implantable cardioverter/defibrillator therapies for AT/AF were activated and de-activated every 3 months in a cross-over study design. The mean follow-up was 18.5 +/- 7.7 months (6.29 +/- 2.2 cross-over/patient). Overall, there were 684 episodes of ventricular tachyarrhythmias in 48.4% of patients (n = 59). In 33.6% of patients (n = 41), 532 supraventricular tachyarrhythmias occurred. Activation of ICD therapies for AT/AF did not result in a reduction of ventricular tachyarrhythmias (P = 0.92). Patients with monomorphic ventricular tachycardias (mVTs) as index arrhythmia for ICD implantation or inducible mVTs in the electrophysiological study had the highest probability of recurrences of ventricular tachyarrhythmias.

CONCLUSION

For patients with standard indications for ICD therapy and no indication for cardiac pacing, a dual-chamber ICD capable of atrial tachyarrhythmia treatment offers no benefit concerning a reduction of ventricular tachyarrhythmias.

Для цитирования: Gradaus Rainer, Seidl Karlheinz, Korte Thomas, Himmrich Ewald, Wieneke Heinrich, Schuchert Andreas, Bauer Wolfgang, Gerss Joachim, Wollmann Christian G, Borggrefe Martin, Böcker Dirk. Reduction of ventricular tachyarrhythmia by treatment of atrial fibrillation in ICD patients with dual-chamber implantable cardioverter/defibrillators capable of atrial therapy delivery: the REVERT-AF Study.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2007. DOI: 10.1093/europace/eum035.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.