Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2006

Use of a new cardiac pacing mode designed to eliminate unnecessary ventricular pacing.

Fröhlig Gerd, Gras Daniel, Victor Jacques, Mabo Philippe, Galley Daniel, Savouré Arnaud, Jauvert Gaël, Defaye Pascal, Ducloux Pascale, Amblard Amel

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

AIMS: To examine the performance of AAIsafeR2, a new pacing mode to minimize the cumulative proportion of ventricular pacing in patients who do not need regular ventricular support.

METHODS AND

RESULTS

The safety of AAIsafeR2 was examined in 123 recipients (73 +/- 12 years old, 51% men) of dual chamber pacemakers implanted for sinus node dysfunction, paroxysmal AV block or the bradycardia-tachycardia syndrome. Data were collected from pacemaker diagnostics, and the first 43 patients underwent 24-h Holter recordings before being discharged from the hospital with AAIsafeR2 activated. No adverse event related to AAIsafeR2 was observed. All ventricular pauses detected on Holter tapes triggered immediate back-up ventricular pacing. Appropriate switches to DDD occurred in 97 of 123 patients. In 69 of 123 devices (56%) switches to DDD were non-sustained, and the average % ventricular pacing in this group was 0.2+/-0.5%.

CONCLUSION

AAIsafeR2 mode seems to be safe and reliable in patients with infrequent slowing or pauses in ventricular activity, while maintaining ventricular pacing below 1%.

Для цитирования: Fröhlig Gerd, Gras Daniel, Victor Jacques, Mabo Philippe, Galley Daniel, Savouré Arnaud, Jauvert Gaël, Defaye Pascal, Ducloux Pascale, Amblard Amel. Use of a new cardiac pacing mode designed to eliminate unnecessary ventricular pacing.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2006. DOI: 10.1093/europace/euj024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.