A novel pinless augmented reality-based navigation system for total hip arthroplasty in the supine position: a comparative study of acetabular cup angle accuracy.

Kurishima Hiroaki, Ogawa Hiroyuki, Yamada Norikazu, Noro Atsushi, Tomata Yasutake, Tsukada Sachiyuki

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

INTRODUCTION

We developed a novel pinless augmented reality (AR)-based portable navigation system for total hip arthroplasty (THA) in the supine position to eliminate pin-site complications and intraoperative navigation abandonment, and to reduce registration time. This study compared the measurement accuracy of this new pinless portable navigation system with that of a conventional AR-based portable navigation system requiring pin insertion.

MATERIALS AND

METHODS

This retrospective cohort study compared primary unilateral THAs performed using the novel pinless AR-based navigation system and a conventional AR-based navigation system. The primary outcome was the absolute difference between the acetabular cup angles displayed on the navigation screen and those measured on postoperative computed tomography. Secondary outcomes included target achievement rates within 5° and 10°. Pin-site complications, intraoperative navigation abandonment, and dislocations were also recorded.

RESULTS

A total of 228 hips were analyzed (Pinless: n = 115; Conventional: n = 113). While there was no significant difference in the median absolute difference for inclination (2.5° vs 2.4°; p = 0.366), the median absolute difference for anteversion was significantly larger in the pinless group than in the conventional group (2.9° vs 1.9°; p = 0.011). There were no significant differences in the percentages within 5° (80.0% vs 77.9%; p = 0.747) and 10° (99.1% vs 97.3%; p = 0.367). No pin-site complications, intraoperative navigation abandonment, or dislocations occurred in either group.

CONCLUSIONS

Despite a small statistical difference in the absolute anteversion error, the pinless AR-based navigation system demonstrated acceptable measurement accuracy and comparable target achievement rates relative to the conventional system requiring pin insertion in this retrospective cohort. While the new pinless AR-based navigation system eliminates the risk of pin-site complications and intraoperative navigation abandonment, and offers instantaneous registration, further prospective studies are warranted to confirm these findings.

Для цитирования: Kurishima Hiroaki, Ogawa Hiroyuki, Yamada Norikazu, Noro Atsushi, Tomata Yasutake, Tsukada Sachiyuki. A novel pinless augmented reality-based navigation system for total hip arthroplasty in the supine position: a comparative study of acetabular cup angle accuracy.. Archives of orthopaedic and trauma surgery. 2026. DOI: 10.1007/s00402-026-06324-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.