DOI 10.3928/1081597x-20180706-01
Comparing Total Keratometry Measurement on the IOLMaster 700 With Goggin Nomogram Adjusted Anterior Keratometry.
Аннотация
PURPOSE: To compare "total keratometry" measurements with Goggin nomogram adjusted anterior keratometry (GNAK) for toric intraocular lens calculation.
METHODS
Routine biometry provided measured total keratometry values from which astigmatism was derived. The Goggin nomogram was applied to anterior keratometry values on the same eyes to provide nomogram-adjusted anterior keratometry values (GNAK) that estimate total corneal astigmatism. The agreement between total keratometry and GNAK was analyzed.
RESULTS
Overall, in 46 eyes there was no statistically significant difference between median GNAK and total keratometry power values (P = .746). No statistically significant difference remained in against-the-rule and oblique subgroup analyses. Absolute and signed steep axis of astigmatism was statistically significantly different for GNAK and total keratometry in the overall analysis (P < .001 and = .029, respectively) and for against-the-rule and oblique subgroup analyses. The with-the-rule subgroup showed a statistically significant difference in astigmatic power and no significant signed steep axis difference between GNAK and total keratometry.
CONCLUSIONS
Total keratometry appears able to measure total corneal astigmatism to match closely (clinically and statistically) GNAK estimation of that parameter. This indicates that it would be safe and reasonable to use total keratometry data for planning of cataract surgery with toric IOLs. [J Refract Surg. 2018;34(8):521-526.].
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.