DOI 10.1016/j.jcrs.2018.04.035
Astigmatism induced by intraocular lens tilt evaluated via ray tracing.
Аннотация
PURPOSE: To evaluate astigmatism induced by aspheric and toric intraocular lens (IOL) tilt using a ray-tracing model.
SETTING
Cullen Eye Institute, Baylor College of Medicine, Houston, Texas, USA.
DESIGN
Experimental study.
METHODS
Ray-tracing eye models with aspheric IOLs (16.0 diopters [D], 22.0 D, and 28.0 D) and toric IOLs (16.0 D, 22.0 D, and 28.0 D each with toricities of 1.50 D, 3.75 D, and 6.00 D) were used. The IOLs were tilted from 1 to 10 degrees horizontally around a 90-degree vertical meridian. Toric IOLs were aligned at 90 degrees and 180 degrees to correct with-the-rule (WTR) and against-the-rule (ATR) corneal astigmatism, respectively. Astigmatism at the corneal plane induced by IOL tilt was calculated.
RESULTS
Induced astigmatism increased with increasing IOL tilt and power. Horizontal tilt around a vertical meridian induced ATR astigmatism. For 5 degrees of tilt, induced astigmatism was 0.08 D, 0.11 D, and 0.14 D for 16.0 D, 22.0 D, and 28.0 D aspheric IOLs, respectively. Ten degrees of IOL tilt produced 0.33 D, 0.44 D, and 0.56 D of induced astigmatism for 16.0 D, 22.0 D, and 28.0 D aspheric IOLs, respectively. Tilting toric IOLs aligned at 90 degrees around a vertical meridian increased the magnitude of induced ATR astigmatism. Tilting toric IOLs aligned at 180 degrees decreased the magnitude of induced WTR astigmatism.
CONCLUSIONS
Tilting aspheric IOLs horizontally around a vertical meridian induced ATR astigmatism. Tilting toric IOLs aligned at 90 degrees increased ATR astigmatism, resulting in overcorrection. Tilting toric IOLs aligned at 180 degrees decreased WTR astigmatism, producing undercorrection.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.