MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/aor.12274

Медицинская библиотека · Artificial organs · 2014

Possible benefits of catheters with lateral holes in coronary thrombus aspiration: a computational study for different clot viscosities and vacuum pressures.

Soleimani Sajjad, Dubini Gabriele, Pennati Giancarlo

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

According to a number of clinical studies, coronary aspiration catheters are useful tools to remove a thrombus (blood clot) blocking a coronary artery. However, these thrombectomy devices may fail to remove the blood clot entirely. Few studies have been devoted to a systematic analysis of factors affecting clot aspiration. The geometric characteristics of the aspiration catheter, the physical properties of the thrombus, and the applied vacuum pressure are crucial parameters. In this study, the aspiration of a blood clot blocking a coronary bifurcation is computationally simulated. The clot is modeled as a highly viscous fluid, and a two-phase (blood and clot) problem is solved. The effects of geometric variations in the tip of the coronary catheter, including lateral hole size and location, are investigated considering different aspiration pressures and clot viscosities. A Bird-Carreau model is adopted for blood viscosity, while a power law model is used to describe the clot rheology. Computational results for blood clot aspiration show that the presence of holes in the lateral part of the tip of the catheter can be beneficial depending on clot viscosity, hole features, and applied aspiration pressure. In general, the holes are beneficial when the clot viscosity is low, while aspiration catheters without any extra lateral holes exhibit better performance for higher clot viscosity. However, when higher aspiration pressures are applied, the catheters tend to behave relatively similarly in removing clots with various viscosities, reducing the role of the clot viscosity.

Для цитирования: Soleimani Sajjad, Dubini Gabriele, Pennati Giancarlo. Possible benefits of catheters with lateral holes in coronary thrombus aspiration: a computational study for different clot viscosities and vacuum pressures.. Artificial organs. 2014. DOI: 10.1111/aor.12274.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.