Low settings of the ventricular pacing output in patients dependent on a pacemaker: are they really safe?

Schuchert Andreas, Frese Jens, Stammwitz Ekkehard, Novák Miroslav, Schleich Arthur, Wagner Stefan M, Meinertz Thomas

Клиническое исследованиеСтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

It is generally acknowledged that pacemaker output must be adjusted with a 100% voltage safety margin above the pacing threshold to avoid ineffective pacing, especially in patients dependent on pacemakers.

AIMS: The aim of this prospective crossover study was to assess the beat-to-beat safety of low outputs in patients who are dependent on a pacemaker between 2 follow-up examinations.

METHODS

The study included 12 patients who had received a DDD pacemaker with an automatic beat-to-beat capture verification function. The ventricular output at 0.4 milliseconds pulse duration was programmed independently of the actual pacing threshold in a crossover randomization to 1.0 V, 1.5 V, and 2.5 V for 6 weeks each. At each follow-up, the diagnostic counters were interrogated and the pacing threshold at 0.4 milliseconds was determined in 0.1-V steps. The diagnostic pacemaker counters depict the frequency of back-up pulses delivered because of a loss of capture. During the randomization to 1.0-V output, we evaluated whether the adjustment of the output under consideration of the >100% voltage safety margin reduced the frequency of back-up pulses.

RESULTS

Pacing thresholds at the randomization to 1.0-V, 1.5-V, and 2.5-V output were not significantly different, with 0.7 +/- 0.3 V at 2.5-V output, 0.6 +/- 0.2 V at 1.5-V output, and 0.6 +/- 0.2 V at 1.0-V output. The frequency of back-up pulses was similar at 2.5-V and 1.5-V output, 2.2% +/- 1.9% and 2.0% +/- 2.0%, respectively. The frequency of back-up pulses significantly increased at 1.0-V output to 5.8% +/- 6.4% (P <.05). Back-up pulses >5% of the time between the 2 follow-ups were observed in no patient at 2.5 V, in 1 patient at 1.5 V, and in 5 patients at 1.0 V. At the randomization to the 1.0-V output, 6 patients had pacing thresholds of 0.5 V or less, and 6 patients had pacing thresholds >0.5 V. The frequency of back-up pulses in the 2 groups was not significantly different, 6.4% +/- 8.6% and 5.7% +/- 2.6%.

CONCLUSIONS

The frequency of back-up pulses was significantly higher at 1.0-V output than at 1.5-V and 2.5-V output. This also applied to patients with pacing thresholds of < or =0.5 V. Fixed low outputs seem not to be absolutely safe between 2 follow-ups in patients who are dependent on a pacemaker, even when the output has a 100% voltage safety margin above the pacing threshold. When patients with pacemakers programmed to a low ventricular output have symptoms of ineffective pacing, an intermittent increase of the pacing threshold should be carefully ruled out.

Для цитирования: Schuchert Andreas, Frese Jens, Stammwitz Ekkehard, Novák Miroslav, Schleich Arthur, Wagner Stefan M, Meinertz Thomas. Low settings of the ventricular pacing output in patients dependent on a pacemaker: are they really safe?. American heart journal. 2002. DOI: 10.1067/mhj.2002.123141.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.