[Comparative studies with various corneal topography measurement devices].
Аннотация
PURPOSE: It is difficult to measure the topography of the cornea with high resolution and visualize it on a map displaying refraction. This is demonstrated by the ongoing improvement and further development of different methods and by the fact that users of these techniques are not always satisfied.
MATERIALS AND
METHODS
Five different ring projectors were compared. Spherical and aspherical ball standards were used to measure the standard deviation of the refraction of the system. A patient group of eight people with healthy eyes was used to compare the measurement accuracy and operational errors. Some patients were also measured after cataract surgery. The lateral and axial range of the devices was determined.
RESULTS
The measurement accuracy for ball standards for the five devices was below 1/8 D. For the in vivo case in the control group it was below 1/4 D for four devices. Furthermore, the reproducibility of the results, the accuracy of angle determination and the influence of mechanical and optical design will be discussed. The study demonstrates applications and limits of this measurement method.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.