m+Medelevator.RU
← Поиск публикаций
Nan fang yi ke da xue xue bao = Journal of Southern Medical University · 2014

[Inter- and intra-observer reproducibility of 3.0 T high-resolution magnetic resonance imaging for evaluating atherosclerotic stenosis in the middle cerebral artery].

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Jia Zejun, Yuan Yuan, Yang Zhigang, Zhao Rui, Xu Yi, Huang Qinghai, Zhao Wenyuan, Ma Xiaolong, Deng Xiaoqun, Hong Bo, Liu Jian-Min

Аннотация

OBJECTIVE: To assess the reproducibility of 3.0 T high-resolution magnetic resonance imaging (HR MRI) for evaluation of atherosclerotic stenosis in the middle cerebral artery (MCA).

METHODS: From February, 2011 to December, 2013, 66 consecutive patients with MCA-M1 atherosclerotic stenosis (50%-99%) confirmed by digital subtractive angiography (DSA) received examinations with 3.0 T HR MRI for measurement of the vessel area (VA) and lumen area (LA) at the maximum narrow site (VA(narrow) and LA(narrow)) and the reference site (VA(reference) and LA(reference)) as well as the plaque distribution (ventral, dorsal, superior, and inferior). Two independent readers reviewed all the images and one reader reevaluated these images 4 weeks later. The inter- and intra-observer reproducibility was evaluated using the intraclass correlation coefficient (ICC).

RESULTS: The measurements of VA(narrow), VA(reference), and LA(reference) using HR MRI showed excellent inter- (ICC=0.801, 0.843, and 0.808, respectively) and intra-observer reproducibility (ICC=0.811, 0.916, and 0.958, respectively), but the measurement of LA(narrow) had only moderate inter- and intra-observer reproducibility (ICC=0.584 and 0.625, respectively). For plaque distribution analysis (ventral, dorsal, superior, and inferior plaques), HR MRI also showed excellent inter- (ICC=0.856, 0.836, 0.791, and 0.905, respectively) and intra-observer reproducibility (ICC=0.876, 0.827, 0.825, and 0.950, respectively).

CONCLUSION: HR MRI shows good inter- and intra-observer reproducibility in identifying MCA-M1 atherosclerotic plaque distribution and vessel and lumen measurements, but its reliability for lumen area measurement at the maximum narrowing site needs to be improved.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты