Results of a multicenter clinical study evaluating a new Smart algorithm to measure neural response imaging.

Akin Istemihan, Götze Romy, Basta Dietmar, Böhnke Britta, Müller-Deile Joachim, Ormezzano Yves, Frachet Bruno, Gérard Jean-Marc, Deggouj Naïma, Philips Birgit, Dhooge Ingeborg, Arnold Laure

Сравнительное исследованиеОценочное исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVE

To evaluate the smart algorithm in speed and reliability of threshold estimation compared with the algorithm available in the standard fitting software and to evaluate the possibility of using programs based on the smart algorithm instead of programs derived from behavioral measures.

PATIENTS: Twenty subjects unilaterally implanted with a CII Bionic Ear or HiRes90K device.

INTERVENTIONS: Neural response imaging thresholds (tNRI) were measured using both the smart approach within the Research Studies Platform for Objective Measures and the SoundWave fitting software. Measurements were performed intraoperatively, at first fitting, and after 3 months of implant use. Each subject received a standard behavioral program and a SmartNRI program. Speech perception tests were conducted at 3 months, and subjective preferences were documented.

MAIN OUTCOME MEASURES: Smart tNRI and SoundWave tNRI at each session; speech test results and subject preferences at the 3-month session.

RESULTS

High correlations were found between smart tNRI and SoundWave tNRI. The time required to obtain NRI thresholds with the smart algorithm was a quarter of the time needed with SoundWave. Although most tested subjects preferred their behavioral programs, there were no significant differences in performance between SmartNRI and behavioral programs.

CONCLUSION

Neural response imaging thresholds were obtained more rapidly with the smart algorithm than with SoundWave. Because no differences were observed between SmartNRI and behavioral programs, SmartNRI programs may be a useful alternative to behavioral programs in difficult to fit cases, where user feedback is sometimes difficult to obtain.

Для цитирования: Akin Istemihan, Götze Romy, Basta Dietmar, Böhnke Britta, Müller-Deile Joachim, Ormezzano Yves, Frachet Bruno, Gérard Jean-Marc, Deggouj Naïma, Philips Birgit, Dhooge Ingeborg, Arnold Laure. Results of a multicenter clinical study evaluating a new Smart algorithm to measure neural response imaging.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2012. DOI: 10.1097/mao.0b013e318254503e.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.