Maxin Anthony J, Kush Sophie, Gulek Bernice G, Winston Graham M, Chae John, Shaibani Rami, McGrath Lynn B, Abecassis Isaac J, Levitt Michael R
Аннотация
OBJECTIVES: Vasospasm is a complication of aneurysmal subarachnoid hemorrhage (aSAH) that can change the trajectory of recovery and is associated with morbidity and mortality. Earlier detection of vasospasm could improve patient outcomes. Our objective is to evaluate the accuracy of smartphone-based quantitative pupillometry in the detection of radiographic vasospasm and delayed cerebral ischemia (DCI) after aSAH.
MATERIALS AND METHODS: We prospectively collected pupillary light reflex (PLR) parameters from patients with aSAH admitted to a neurocritical care unit at a single hospital twice daily using quantitative smartphone pupillometry recordings. PLR parameters included: Maximum pupil diameter, minimum pupil diameter, percent change in pupil diameter, latency in beginning of pupil constriction to light, mean constriction velocity, maximum constriction velocity, and mean dilation velocity. Two-tailed t-tests for independent samples were performed to determine changes in average concurrent PLR parameter values between the following comparisons: (1) patients with and without radiographic vasospasm (defined by angiography with the need for endovascular intervention) and (2) patients with and without DCI.
RESULTS: 49 subjects with aSAH underwent 323 total PLR recordings. For PLR recordings taken with (n=35) and without (n=241) radiographic vasospasm, significant differences were observed in MIN (35.0 ± 7.5 pixels with vasospasm versus 31.6 ± 6.2 pixels without; p=0.002). For PLR recordings taken with (n=43) and without (n=241) DCI, significant differences were observed in MAX (48.9 ± 14.3 pixels with DCI versus 42.5 ± 9.2 pixels without; p<0.001).
CONCLUSIONS: Quantitative smartphone pupillometry has the potential to be used to detect radiographic vasospasm and DCI after aSAH.
MATERIALS AND METHODS: We prospectively collected pupillary light reflex (PLR) parameters from patients with aSAH admitted to a neurocritical care unit at a single hospital twice daily using quantitative smartphone pupillometry recordings. PLR parameters included: Maximum pupil diameter, minimum pupil diameter, percent change in pupil diameter, latency in beginning of pupil constriction to light, mean constriction velocity, maximum constriction velocity, and mean dilation velocity. Two-tailed t-tests for independent samples were performed to determine changes in average concurrent PLR parameter values between the following comparisons: (1) patients with and without radiographic vasospasm (defined by angiography with the need for endovascular intervention) and (2) patients with and without DCI.
RESULTS: 49 subjects with aSAH underwent 323 total PLR recordings. For PLR recordings taken with (n=35) and without (n=241) radiographic vasospasm, significant differences were observed in MIN (35.0 ± 7.5 pixels with vasospasm versus 31.6 ± 6.2 pixels without; p=0.002). For PLR recordings taken with (n=43) and without (n=241) DCI, significant differences were observed in MAX (48.9 ± 14.3 pixels with DCI versus 42.5 ± 9.2 pixels without; p<0.001).
CONCLUSIONS: Quantitative smartphone pupillometry has the potential to be used to detect radiographic vasospasm and DCI after aSAH.