MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.parkreldis.2017.10.015

Медицинская библиотека · Parkinsonism & related disorders · 2017

Square biphasic pulse deep brain stimulation for essential tremor: The BiP tremor study.

De Jesus Sol, Almeida Leonardo, Shahgholi Leili, Martinez-Ramirez Daniel, Roper Jaimie, Hass Chris J, Akbar Umer, Wagle Shukla Aparna, Raike Robert S, Okun Michael S

Сравнительное исследованиеСтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Conventional deep brain stimulation (DBS) utilizes regular, high frequency pulses to treat medication-refractory symptoms in essential tremor (ET). Modifications of DBS pulse shape to achieve improved effectiveness is a promising approach.

OBJECTIVES

The current study assessed the safety, tolerability and effectiveness of square biphasic pulse shaping as an alternative to conventional ET DBS.

METHODS

This pilot study compared biphasic pulses (BiP) versus conventional DBS pulses (ClinDBS). Eleven ET subjects with clinically optimized ventralis intermedius nucleus DBS were enrolled. Objective measures were obtained over 3 h while ON BiP stimulation.

RESULTS

There was observed benefit in the Fahn-Tolosa Tremor Rating Scale (TRS) for BiP conditions when compared to the DBS off condition and to ClinDBS setting. Total TRS scores during the DBS OFF condition (28.5 IQR = 24.5-35.25) were significantly higher than the other time points. Following active DBS, TRS improved to (20 IQR = 13.8-24.3) at ClinDBS setting and to (16.5 IQR = 12-20.75) at the 3 h period ON BiP stimulation (p = 0.001). Accelerometer recordings revealed improvement in tremor at rest (χ2 = 16.1, p = 0.006), posture (χ2 = 15.9, p = 0.007) and with action (χ2 = 32.1, p=<0.001) when comparing median total scores at ClinDBS and OFF DBS conditions to 3 h ON BiP stimulation. There were no adverse effects and gait was not impacted.

CONCLUSION

BiP was safe, tolerable and effective on the tremor symptoms when tested up to 3 h. This study demonstrated the feasibility of applying a novel DBS waveform in the clinic setting. Larger prospective studies with longer clinical follow-up will be required.

Для цитирования: De Jesus Sol, Almeida Leonardo, Shahgholi Leili, Martinez-Ramirez Daniel, Roper Jaimie, Hass Chris J, Akbar Umer, Wagle Shukla Aparna, Raike Robert S, Okun Michael S. Square biphasic pulse deep brain stimulation for essential tremor: The BiP tremor study.. Parkinsonism & related disorders. 2017. DOI: 10.1016/j.parkreldis.2017.10.015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.