DOI 10.1016/j.parkreldis.2017.03.017
Thalamic DBS with a constant-current device in essential tremor: A controlled clinical trial.
Аннотация
INTRODUCTION
This study of thalamic deep brain stimulation (DBS) investigated whether a novel constant-current device improves tremor and activities of daily living (ADL) in patients with essential tremor (ET).
METHODS
A prospective, controlled, multicenter study was conducted at 12 academic centers. We investigated the safety and efficacy of unilateral and bilateral constant-current DBS of the ventralis intermedius (VIM) nucleus of the thalamus in patients with essential tremor whose tremor was inadequately controlled by medications. The primary outcome measure was a rater-blinded assessment of the change in the target limb tremor score in the stimulation-on versus stimulation-off state six months following surgery. Multiple secondary outcomes were assessed at one-year follow-up, including motor, mood, and quality-of-life measures.
RESULTS
127 patients were implanted with VIM DBS. The blinded, primary outcome variable (n = 76) revealed a mean improvement of 1.25 ± 1.26 points in the target limb tremor rating scale (TRS) score in the arm contralateral to DBS (p < 0.001). Secondary outcome variables at one year revealed significant improvements (p ≤ 0.001) in quality of life, depression symptoms, and ADL scores. Forty-seven patients had a second contralateral VIM-DBS, and this group demonstrated reduction in second-sided tremor at 180 days (p < 0.001). Serious adverse events related to the surgery included infection (n = 3), intracranial hemorrhage (n = 3), and device explantation (n = 3).
CONCLUSION
Unilateral and bilateral constant-current VIM DBS significantly improves upper extremity tremor, ADL, quality of life, and depression in patients with severe ET.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.