DOI 10.1007/s00701-017-3090-7

Медицинская библиотека · Acta neurochirurgica · 2017

Rapid battery depletion and loss of therapy due to a short circuit in bipolar DBS for essential tremor.

Allert Niels, Barbe Michael Thomas, Timmermann Lars, Coenen Volker Arnd

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

Technical dysfunctions have been reported reducing efficacy of deep brain stimulation (DBS). Here, we report on an essential-tremor patient in whom a short circuit in bipolar DBS resulted not only in unilateral loss of therapy but also in high current flow and thereby rapid decline of the impulse-generator battery voltage from 2.83 V a week before the event to 2.54 V, indicating the need for an impulse-generator replacement. Immediate re-programming restored therapeutic efficacy. Moreover, the reduction in current flow allowed the battery voltage to recover without immediate surgical intervention to 2.81 V a week later.

Для цитирования: Allert Niels, Barbe Michael Thomas, Timmermann Lars, Coenen Volker Arnd. Rapid battery depletion and loss of therapy due to a short circuit in bipolar DBS for essential tremor.. Acta neurochirurgica. 2017. DOI: 10.1007/s00701-017-3090-7.