Method evaluation of pepsinogen I/II assay based on chemiluminescent immunoassays and comparison with other test methods.
Аннотация
BACKGROUND
Serum pepsinogen (PG) I and the PG I/PG II ratio have been used for atrophic gastritis (AG) diagnosis for decades. Low levels of PG I and/or PG I/PG II are closely related to AG and predict the risk of gastric cancer. We evaluated the performance of the chemiluminescent immunoassay-based Architect Pepsinogen I/II assay.
METHODS
The evaluation consisted of determination of the precision, linearity, limit of blank (LoB), limit of detection (LoD) and method comparison with Eiken and Biohit assays.
RESULTS
The total CVs were below 5% for both PG I and PG II. Acceptable linearity was observed for PG I and PG II in their respective reportable ranges. The PG I LoB was 0.317ng/mL and the PG II LoB was 0.418ng/mL, and LoDs were 0.412ng/mL and 0.497ng/mL, respectively. Correlation analysis indicated that results of the Architect assay were comparable to those of the Eiken and Biohit assays, but the three methods lead to different estimations of the cancer risk.
CONCLUSION
The overall analytical performance of Architect Pepsinogen I/II assay is acceptable for the detection of patients with suspected AG. The categorization results of gastric cancer risk showed some difference among test methods suggesting the need for harmonization among the methods from vendors.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.