A Case of a Cardiac Resynchronization Therapy-Defibrillator Exhibiting a Lower and Alternately Variable Basic Rate.

Клинический случайСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Iwazaki Keigo, Kojima Toshiya, Murasawa Takahide, Yokota Jun, Tanimoto Hikaru, Matsuda Jun, Fukuma Nobuaki, Matsubara Takumi, Shimizu Yu, Oguri Gaku, Hasumi Eriko, Kubo Hitoshi, Chang Kyungho, Fujiu Katsuhito, Komuro Issei

Аннотация

A cardiac resynchronization therapy defibrillator (CRT-D) (Medtronic Inc. Protecta XT) was implanted in a 67-year-old man who had cardiac sarcoidosis with extremely low cardiac function. He had ventricular tachycardia which was controlled by catheter ablation, medication and pacing. The programmed mode was DDI, lower rate was 90 beats/minute, paced AV delay was 150 ms, and the noncompetitive atrial pacing (NCAP) function was programmed as 300 ms.After his admission for pneumonia and heart failure, we changed his DDI mode to a DDD mode because he had atrial tachycardia, which led to inadequate bi-ventricular pacing. After a while, there were cycle lengths which were longer than his device setting and alternately varied. We were able to avoid this phenomenon with AV delay of 120 ms and NCAP of 200 ms.NCAP is an algorithm which creates a gap above a certain period after the detection of an atrial signal during the postventricular atrial refractory period of the pacemaker. This is to prevent atrial tachycardia and repetitive non-reentrant ventriculoatrial (VA) synchrony in the presence of retrograde VA conduction. But in this case, NCAP algorithm induced much lower rate than the programmed basic lower rate. This situation produced some arrhythmias and exacerbated symptoms of heart failure. This had to be paid attention to, especially when the device was programmed at high basic heart rate.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты