DOI 10.1007/s10334-026-01328-9
Standardizing quantitative muscle MRI: an inter-vendor evaluation in healthy adults.
Аннотация
OBJECTIVES
Quantitative muscle MRI (qMRI) is applied to detect muscular alterations, particularly in neuromuscular diseases (NMDs). As studies on rare NMDs benefit from multi-center designs, inter-vendor reproducibility must be assessed to ensure data comparability. However, manufacturer-related variability has been limited investigated. This study evaluates inter-vendor reproducibility of qMRI parameters in healthy subjects.
METHODS
Ten healthy volunteers were investigated within 1 week on two 3T MRI systems (Siemens, Philips). Twelve leg muscles were analyzed using Dixon (fat fraction, FF), T2 mapping (water T2, wT2), and diffusion-weighted-sequence (diffusion parameters: fractional anisotropy (FA), mean, radial, axial diffusivity [MD, RD, λ1]). Reproducibility was assessed using intraclass correlation coefficient (ICC), within-subject coefficients of variation, t tests, and Bland-Altman analysis.
RESULTS
ICCs demonstrated moderate to excellent reproducibility for most parameters (ICCFF = 0.923, ICCwT2 = 0.792, ICCFA = 0.777, ICCMD = 0.708, ICCRD = 0.850), except for λ1 (ICC = 0.361). Significant inter-vendor differences were observed for all parameters (p < 0.05), with systematic biases ranging from 44.2% for FF to 3.6% for FA.
DISCUSSION: qMRI parameters were generally reproducible in healthy volunteers. However, absolute values differed between manufacturers and require correction. These differences should be considered to minimize manufacturer-related discrepancies and to facilitate data pooling in multi-center and longitudinal studies.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.