DOI 10.1016/j.ejrad.2007.10.027
Thin-section CT of lung without ECG gating: 64-detector row CT can markedly reduce cardiac motion artifact which can simulate lung lesions.
Аннотация
PURPOSE: Motion artifacts, which can mimic thickened bronchial wall and the cystic appearance of bronchiectasis, constitute a potential pitfall in the diagnosis of interstitial or bronchial disease. Therefore, purpose of our study was to evaluate whether 64-detector row CT (64-MDCT) enables a reduction in respiratory or cardiac motion artifacts in the lung area on thin-section CT without ECG gating, and to examine the correlation between cardiac motion artifact and heart rate.
MATERIALS AND
METHODS
Thirty-two patients with suspected diffuse lung disease, who underwent both 8- and 64-MDCT (gantry rotation time, 0.5 and 0.4s, respectively), were included. The heart rates of an additional 155 patients were measured (range, 48-126 beats per minute; mean, 76 beats per minute) immediately prior to 64-MDCT, and compared to the degree of cardiac motion artifact. Two independent observers evaluated the following artifacts on a monitor without the knowledge of relevant clinical information: (1) artifacts on 8- and 64-MDCT images with 1.25-mm thickness and those on 64-MDCT images with 0.625-mm thickness in 32 patients; and (2) artifacts on 64-MDCT images with 0.625-mm thickness in 155 patients.
RESULTS
Interobserver agreement was good in evaluating artifacts on 8-MDCT images with 1.25-mm thickness (weighted Kappa test, kappa=0.61-0.71), and fair or poor in the other evaluations (kappa<0.31). Two observers stated that cardiac motion artifacts were more significant on 8-MDCT than on 64-MDCT in all 32 patients. Statistically significant differences were found at various checkpoints only in comparing artifacts between 8- and 64-MDCT for 1.25-mm thickness (Wilcoxon's signed-rank test, p<0.0017). Cardiac motion artifacts on 64-MDCT had no significant correlation with heart rate (Spearman's correlation coefficient by rank test).
CONCLUSION
The high temporal resolution of 64-MDCT appears to reduce cardiac motion artifact that can affect thin-section scans of the lung parenchyma.