Pacing and clinical electrophysiology : PACE · 2023
Catheter ablation of atrial fibrillation using FireMagic TrueForce ablation catheter: The TRUEFORCE trial.
Многоцентровое исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Zhang Rongfeng, Chu Huimin, Liu Shaowen, Yang Bing, Han Bing, Xiao Xianjie, Ma Chengming, Sun Yuanjun, Yin Xiaomeng, Xia Yunlong, Gao Lianjun, TRUEFORCE investigators.
Аннотация
BACKGROUND: The use of contact force (CF) sensing catheters has provided a revolutionary improvement in catheter ablation (CA) of atrial fibrillation (AF) in the past decade. However, the success rate of CA for AF remains limited, and some complications still occur.
METHODS: The TRUEFORCE trial (Catheter Ablation of Atrial Fibrillation using FireMagic TrueForce Ablation Catheter) is a multicenter, prospective, single-arm objective performance criteria study of AF patients who underwent their first CA procedure using FireMagic TrueForce ablation catheter.
RESULTS: A total of 120 patients (118 with paroxysmal AF) were included in this study, and 112 patients included in the per-protocol analysis. Pulmonary vein isolation (PVI) was achieved in 100% of the patients, with procedure and fluoroscopy time of 146.63 ± 40.51 min and 12.89 ± 5.59 min, respectively. Freedom from recurrent atrial arrhythmia after ablation was present 81.25% (95% confidence interval [CI]: 72.78%-88.00%) of patients. No severe adverse events (death, stroke/transient ischemic attack [TIA], esophageal fistula, myocardial infarction, thromboembolism, or pulmonary vein stenosis) were detected during the follow-up. Four (4/115, 3.33%) adverse events were documented, including one abdominal discomfort, one femoral artery hematoma, one coughing up blood, and one postoperative palpitation and insomnia.
CONCLUSIONS: This study demonstrated the clinical feasibility of FireMagic force-sensing ablation catheter in CA of AF, with a satisfactory short- and long-term efficacy and safety.
METHODS: The TRUEFORCE trial (Catheter Ablation of Atrial Fibrillation using FireMagic TrueForce Ablation Catheter) is a multicenter, prospective, single-arm objective performance criteria study of AF patients who underwent their first CA procedure using FireMagic TrueForce ablation catheter.
RESULTS: A total of 120 patients (118 with paroxysmal AF) were included in this study, and 112 patients included in the per-protocol analysis. Pulmonary vein isolation (PVI) was achieved in 100% of the patients, with procedure and fluoroscopy time of 146.63 ± 40.51 min and 12.89 ± 5.59 min, respectively. Freedom from recurrent atrial arrhythmia after ablation was present 81.25% (95% confidence interval [CI]: 72.78%-88.00%) of patients. No severe adverse events (death, stroke/transient ischemic attack [TIA], esophageal fistula, myocardial infarction, thromboembolism, or pulmonary vein stenosis) were detected during the follow-up. Four (4/115, 3.33%) adverse events were documented, including one abdominal discomfort, one femoral artery hematoma, one coughing up blood, and one postoperative palpitation and insomnia.