Programming a fixed long atrioventricular delay is not effective in preventing ventricular pacing in patients with sick sinus syndrome.

Nielsen J C, Pedersen A K, Mortensen P T, Andersen H R

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

AIM: Most patients with sick sinus syndrome (SSS) and normal atrioventricular (AV) conduction receive dual chamber (DDDR) pacemakers. Programming a long AV delay has been proposed to avoid ventricular pacing. The present study aimed to evaluate ventricular stimulation in SSS patients with DDDR pacemakers with a long AV delay.

METHODS AND

RESULTS

Thirty eight patients treated with DDDR pacemakers with a fixed long AV delay (300 ms) were studied. Data from the pacemaker event recorder were retrieved after 3 months and every year after implantation. Ten patients underwent 24 h Holter recording. Mean follow-up was 11.9 +/- 8.3 months. Median daily number of paced events in the ventricle was 2659 (25th-75th percentiles: 775-21 315) with a large inter-individual variation. The proportion of paced events in the ventricle correlated weakly with the baseline PQ interval (Spearman's rho 0.331, P = 0.043). In 12/38 patients the mean daily number of paced events in the ventricle exceeded 10,000. During 24 h Holter recording, pacemaker arrhythmias caused by repetitive retrograde atrioventricular conduction, known as VA (ventriculoatrial) conduction, occurred in five out of 10 patients.

CONCLUSION

DDDR pacing with a fixed long AV delay is inefficient in reducing ventricular pacing in one third of patients and is associated with a high risk of arrhythmias caused by repetitive retrograde AV conduction, and therefore cannot be recommended for general use in SSS patients.

Для цитирования: Nielsen J C, Pedersen A K, Mortensen P T, Andersen H R. Programming a fixed long atrioventricular delay is not effective in preventing ventricular pacing in patients with sick sinus syndrome.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 1999. DOI: 10.1053/eupc.1998.0026.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.