Anomalous "VVI" pacing of a dual chamber pacemaker when programmed in DDI(R) mode.

Masini M, Dianda R, Cortigiani L, Nannini E

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

We describe the case of a dual chamber rate responsive pacemaker (Relay, model 294-03, Intermedics, Angleton, TX, USA) implanted in a 68-year-old male for sick sinus syndrome, which was not working properly when programmed in the DDIR mode, thus determining occasionally a sort of "VVI" pacing. However, the pacemaker performed well when programmed in the DDDR mode. We discovered that this was not a malfunction of a single device but rather a general behavior of this family of Intermedics dual chamber pacemakers (also not rate responsive), caused by a software problem.

Для цитирования: Masini M, Dianda R, Cortigiani L, Nannini E. Anomalous "VVI" pacing of a dual chamber pacemaker when programmed in DDI(R) mode.. Pacing and clinical electrophysiology : PACE. 1999. DOI: 10.1111/j.1540-8159.1999.tb00637.x.