DOI 10.1007/s10792-017-0602-7

Медицинская библиотека · International ophthalmology · 2017

Validation of Fourier analysis of videokeratographic data.

Sideroudi Haris, Labiris Georgios, Ditzel Fienke, Tsaragli Efi, Georgatzoglou Kimonas, Siganos Haralampos, Kozobolis Vassilios

СтатьяНаблюдательное исследованиеВалидационное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: The aim was to assess the repeatability of Fourier transfom analysis of videokeratographic data using Pentacam in normal (CG), keratoconic (KC) and post-CXL (CXL) corneas.

METHODS

This was a prospective, clinic-based, observational study. One randomly selected eye from all study participants was included in the analysis: 62 normal eyes (CG group), 33 keratoconus eyes (KC group), while 34 eyes, which had already received CXL treatment, formed the CXL group. Fourier analysis of keratometric data were obtained using Pentacam, by two different operators within each of two sessions. Precision, repeatability and Intraclass Correlation Coefficient (ICC), were calculated for evaluating intrassesion and intersession repeatability for the following parameters: Spherical Component (SphRmin, SphEcc), Maximum Decentration (Max Dec), Regular Astigmatism, and Irregularitiy (Irr). Bland-Altman analysis was used for assessing interobserver repeatability.

RESULTS

All parameters were presented to be repeatable, reliable and reproductible in all groups. Best intrasession and intersession repeatability and reliability were detected for parameters SphRmin, SphEcc and Max Dec parameters for both operators using ICC (intrasession: ICC > 98%, intersession: ICC > 94.7%) and within subject standard deviation. Best precision and lowest range of agreement was found for the SphRmin parameter (CG: 0.05, KC: 0.16, and CXL: 0.2) in all groups, while the lowest repeatability, reliability and reproducibility was detected for the Irr parameter.

CONCLUSIONS

The Pentacam system provides accurate measurements of Fourier tranform keratometric data. A single Pentacam scan will be sufficient for most clinical applications.

Для цитирования: Sideroudi Haris, Labiris Georgios, Ditzel Fienke, Tsaragli Efi, Georgatzoglou Kimonas, Siganos Haralampos, Kozobolis Vassilios. Validation of Fourier analysis of videokeratographic data.. International ophthalmology. 2017. DOI: 10.1007/s10792-017-0602-7.