DOI 10.3928/1081597x-20170213-01

Медицинская библиотека · Journal of refractive surgery (Thorofare, N.J. : 1995) · 2017

Biomechanical Characterization of Subclinical Keratoconus Without Topographic or Tomographic Abnormalities.

Vinciguerra Riccardo, Ambrósio Renato, Roberts Cynthia J, Azzolini Claudio, Vinciguerra Paolo

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To present a case series of patients with subclinical keratoconus with normal topometric (anterior curvature) and tomographic findings in one eye who showed abnormalities detected by Corvis ST (Oculus Optikgeräte GmbH, Wetzlar, Germany) in vivo biomechanical assessment.

METHODS

All patients had a complete ophthalmic examination, including the Corvis ST biomechanical measurements, optical tomography, and pachymetry with Pentacam (Oculus Optikgeräte GmbH), and Placido-based topography with either the Nidek (OPD III Nidek, Gamagori, Japan) or CSO platform (Costruzione Strumenti Oftalmici, Florence, Italy). Inclusion criteria were a clinical diagnosis of ectasia in one eye and normal topometric and tomographic findings in the fellow eye (subclinical keratoconus), including a Belin/Ambrósio Enhanced Ectasia total deviation index from the Pentacam with less than 1.6 standard deviations from normative values and a Corvis Biomechanical Index score of greater than 0.5 in both eyes.

RESULTS

Tomographic and topographic analysis was normal in one eye and abnormal in the fellow eye in 12 patients. The biomechanical results with the Corvis Biomechanical Index were shown to be abnormal in both eyes of all patients and aided the diagnosis.

CONCLUSIONS

Biomechanical analysis showed abnormalities, whereas tomography and topography were normal. Basing on these findings, the authors suggest the use of biomechanics as an additional diagnostic tool. [J Refract Surg. 2017;33(6):399-407.].

Для цитирования: Vinciguerra Riccardo, Ambrósio Renato, Roberts Cynthia J, Azzolini Claudio, Vinciguerra Paolo. Biomechanical Characterization of Subclinical Keratoconus Without Topographic or Tomographic Abnormalities.. Journal of refractive surgery (Thorofare, N.J. : 1995). 2017. DOI: 10.3928/1081597x-20170213-01.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.