DOI 10.1136/bjophthalmol-2017-310355

Медицинская библиотека · The British journal of ophthalmology · 2017

Comparison of corneal dynamic parameters and tomographic measurements using Scheimpflug imaging in keratoconus.

Chan Tommy Cy, Wang Yu Meng, Yu Marco, Jhanji Vishal

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

AIM: To compare the diagnostic ability of corneal tomography and corneal dynamic response measurements in normal and keratoconus eyes.

METHODS

Consecutive patients with grade II-III keratoconus and age-matched normal subjects were recruited. Corneal imaging was performed using Pentacam (Oculus Optikgeräte, Wetzlar, Germany) and Corvis (Oculus Optikgeräte). A beta version of Corvis software was used with three additional parameters: maximal change of arc length, deformation amplitude (DA) ratio 1 and DA ratio 2. Diagnostic ability of both devices to differentiate normal and keratoconus eyes was evaluated using receiver-operating characteristic (ROC) curves. The areas under the ROC curve (AUC) and partial AUC (pAUC) for specificity ≥80% for each parameter of Corvis and final D value of Belin/Ambrosio Enhanced Ectasia Display (BAD) were compared.

RESULTS

Forty-two eyes of 42 patients (21 patients with keratoconus and 21 normal subjects) were included. Both groups were age matched (p=0.760). The ROC analysis showed that the final D value of BAD had the highest AUC (0.994) and pAUC (0.194). Maximum inverse radius had the highest AUC (0.954) but a relatively lower pAUC (0.158), while DA ratio 2 had the second highest AUC (0.946) together with the highest pAUC (0.177) among Corvis parameters. There was no significant difference between AUC and pAUC of BAD compared with those of DA ratio 1 (p≥0.162) and DA ratio 2 (p≥0.208).

CONCLUSIONS

The results of our study suggest that Corvis measurements have the potential to differentiate keratoconus and normal eyes. The diagnostic ability of novel parameters on Corvis was comparable to Pentacam.

Для цитирования: Chan Tommy Cy, Wang Yu Meng, Yu Marco, Jhanji Vishal. Comparison of corneal dynamic parameters and tomographic measurements using Scheimpflug imaging in keratoconus.. The British journal of ophthalmology. 2017. DOI: 10.1136/bjophthalmol-2017-310355.