Effect of post crosslinking haze on the repeatability of Scheimpflug-based and slit-scanning imaging devices.
Аннотация
PURPOSE: The aim of this study was to analyze the effect of postcollagen crosslinking (CXL) haze on the measurement and repeatability of pachymetry and mean keratometry (Km) of four corneal topographers.
MATERIALS AND
METHODS
Sixty eyes of sixty patients with progressive keratoconus who had undergone accelerated CXL (ACXL) underwent imaging with a scanning slit imaging device (Orbscan II) and three Scheimpflug imaging devices (Pentacam HR, Sirius, and Galilei). Post-ACXL haze was measured using the densitometry software on the Pentacam HR. Readings of the thinnest corneal thickness (TCT) and Km from three scans of each device were analyzed. Effect of haze on the repeatability of TCT and Km measurements was evaluated using regression models. Repeatability was assessed by coefficient of variation.
RESULTS
Corneal densitometry in different zones affected the repeatability of TCT measurement of Orbscan (P < 0.05) significantly but not the repeatability of TCT with Pentacam HR and Sirius (P = 0.03 and 0.05, respectively). Km values were affected by haze when measured with the Pentacam HR (P < 0.05). The repeatability of Km readings for all devices was unaffected by haze. In the anterior 0-2 mm and 2-6 mm zone, TCT (P = 0.43 and 0.45, respectively), Km values (P = 0.4 and 0.6, respectively), repeatability of TCT (P = 0.1 in both zones), and Km (P = 0.5 and 0.1, respectively) with Galilei were found to be the most reliable.
CONCLUSION
Galilei measurements appear to be least affected by post-ACXL haze when compared with other devices. Hence, topography measurements in the presence of haze need to be interpreted with caution.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.