Contact lens & anterior eye : the journal of the British Contact Lens Association · 2017

Repeatability of topographic corneal thickness in keratoconus comparing Visante™ OCT and Oculus Pentacam HR® topographer.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Otchere Heinz, Sorbara Luigina

Аннотация

PURPOSE: To assess the repeatability and reproducibility of Visante™ OCT and Oculus Pentacam HR® in measuring central (CCT) and topographic corneal thickness (TCT) along the principle meridians in keratoconus (KC) corneae.

METHOD: Twenty participants diagnosed with KC were recruited. There were two study visits. On the first study visit, two repeated corneal thickness measurements were obtained with each instrument. Measurements were repeated at least 48hrs later in the same order. TCT were recorded in the 90, 180, 45 and 135 meridians at 1mm intervals across 8mm chord.

RESULTS: Mean CCT for Visante™ OCT was 484.97±43.14μm (range: 484.84-486.09) and Oculus Pentacam HR® was 478.86±45.31μm (range: 477.20-480.53). No significant difference in TCT between the two visits (p=0.54) and measurements (p=0.63) for Visante™ OCT. For Oculus Pentacam HR®, no significant difference was found for each visit (p=0.18) but differences existed in the measurements outside of the central region (p=0.001). Tukey post-hoc analysis shows the differences (p˂0.05) were found in the +1 and +4 (supero-temporal) locations in the 135 meridian. Significant differences were found comparing the two instruments, (p<0.05). Bland Altman plots were used to demonstrate the differences between the two instruments and indicate their limits of agreement.

CONCLUSION: Both instruments gave repeatable measurements as no significant differences were found in most locations in all meridians. Comparing the two instruments, they were not reproducible in all locations.