MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2196/45760

Медицинская библиотека · Journal of medical Internet research · 2023

Diagnostic Value of a Wearable Continuous Electrocardiogram Monitoring Device (AT-Patch) for New-Onset Atrial Fibrillation in High-Risk Patients: Prospective Cohort Study.

Kwun Ju-Seung, Lee Jang Hoon, Park Bo Eun, Park Jong Sung, Kim Hyeon Jeong, Kim Sun-Hwa, Jeon Ki-Hyun, Cho Hyoung-Won, Kang Si-Hyuck, Lee Wonjae, Youn Tae-Jin, Chae In-Ho, Yoon Chang-Hwan

Многоцентровое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

While conventional electrocardiogram monitoring devices are useful for detecting atrial fibrillation, they have considerable drawbacks, including a short monitoring duration and invasive device implantation. The use of patch-type devices circumvents these drawbacks and has shown comparable diagnostic capability for the early detection of atrial fibrillation.

OBJECTIVE

We aimed to determine whether a patch-type device (AT-Patch) applied to patients with a high risk of new-onset atrial fibrillation defined by the congestive heart failure, hypertension, age ≥75 years, diabetes mellitus, stroke, vascular disease, age 65-74 years, sex scale (CHA2DS2-VASc) score had increased detection rates.

METHODS

In this nonrandomized multicenter prospective cohort study, we enrolled 320 adults aged ≥19 years who had never experienced atrial fibrillation and whose CHA2DS2-VASc score was ≥2. The AT-Patch was attached to each individual for 11 days, and the data were analyzed for arrhythmic events by 2 independent cardiologists.

RESULTS

Atrial fibrillation was detected by the AT-Patch in 3.4% (11/320) of patients, as diagnosed by both cardiologists. Interestingly, when participants with or without atrial fibrillation were compared, a previous history of heart failure was significantly more common in the atrial fibrillation group (n=4/11, 36.4% vs n=16/309, 5.2%, respectively; P=.003). When a CHA2DS2-VASc score ≥4 was combined with previous heart failure, the detection rate was significantly increased to 24.4%. Comparison of the recorded electrocardiogram data revealed that supraventricular and ventricular ectopic rhythms were significantly more frequent in the new-onset atrial fibrillation group compared with nonatrial fibrillation group (3.4% vs 0.4%; P=.001 and 5.2% vs 1.2%; P<.001), respectively.

CONCLUSIONS

This study detected a moderate number of new-onset atrial fibrillations in high-risk patients using the AT-Patch device. Further studies will aim to investigate the value of early detection of atrial fibrillation, particularly in patients with heart failure as a means of reducing adverse clinical outcomes of atrial fibrillation.

TRIAL REGISTRATION: ClinicalTrials.gov NCT04857268; https://classic.clinicaltrials.gov/ct2/show/NCT04857268.

Для цитирования: Kwun Ju-Seung, Lee Jang Hoon, Park Bo Eun, Park Jong Sung, Kim Hyeon Jeong, Kim Sun-Hwa, Jeon Ki-Hyun, Cho Hyoung-Won, Kang Si-Hyuck, Lee Wonjae, Youn Tae-Jin, Chae In-Ho, Yoon Chang-Hwan. Diagnostic Value of a Wearable Continuous Electrocardiogram Monitoring Device (AT-Patch) for New-Onset Atrial Fibrillation in High-Risk Patients: Prospective Cohort Study.. Journal of medical Internet research. 2023. DOI: 10.2196/45760.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.