DOI 10.1016/j.jfo.2016.06.004

Медицинская библиотека · Journal francais d'ophtalmologie · 2016

Reproducibility of single-pass donor DSAEK tissue preparation with the MORIA single-use microkeratome.

Saunier V, Robinet Perrin A, Costet C, Touboul D

Клиническое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To assess the reproducibility of single-pass cutting for preparation of ultra-thin (≤120μm) donor cornea grafts in Descemet stripping automated endothelial keratoplasty (DSAEK).

METHODS

All consecutive patients of DSAEK performed using the MORIA One Use microkeratome (MORIA, Antony, France) in between June 2014 and August 2015. Patient and donor central corneal thickness (CTT), microkeratome head, remaining stromal bed thickness (RSBT), number of cut and graft thickness at 1 month and 6 months postoperatively were recorded in this single-center prospective study.

RESULTS

The mean preoperative donor graft CCT was 569.0±45.1μm and the mean donor endothelial graft CCT immediately after the cut was 116.5±28.7μm. At 1 month postoperatively, the mean CCT was 102.8±35.9μm. At 6 months, the mean CCT was 89.4±26.2μm. In all eyes, the mean CCT decreased from the post-cut (116.5±28.7μm) to the last visit (89.4±26.2μm) (P<0.01) due to in vivo deturgesence of the graft. We obtained 77.5% ultra-thin DSAEK immediately after the cut, 77.5% at 1 month and 89.8% after 6 months.

CONCLUSION

Single-pass technique with the MORIA One Use microkeratome offers safe and reproducible DSAEK tissue preparation and allows achieving ultra-thin DSAEK in 89.8% of cases.

Для цитирования: Saunier V, Robinet Perrin A, Costet C, Touboul D. Reproducibility of single-pass donor DSAEK tissue preparation with the MORIA single-use microkeratome.. Journal francais d'ophtalmologie. 2016. DOI: 10.1016/j.jfo.2016.06.004.