DOI 10.1016/j.jcrs.2016.04.033

Медицинская библиотека · Journal of cataract and refractive surgery · 2016

Agreement of corneal measurements between dual rotating Scheimpflug-Placido system and Placido-based topography device in normal and keratoconus eyes.

Ortiz-Toquero Sara, Zuñiga Virginia, Rodriguez Guadalupe, de Juan Victoria, Martin Raul

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the anterior corneal measurements between Placido-based topography and dual-Scheimpflug topography in healthy and keratoconus eyes.

SETTING

Optometry Research Group, Instituto Universitario de Oftalmobiología Aplicada, University of Valladolid, Valladolid, Spain.

DESIGN

Comparative case series.

METHODS

The mean simulated keratometry (K), flat K, steep K, astigmatism power, corneal astigmatism axis, J0, J45, maximum corneal power point, and white-to-white (WTW) distance were collected and compared between healthy eyes and keratoconus eyes.

RESULTS

The study evaluated in 56 healthy eyes and 56 keratoconus eyes. Placido-based topography underestimated all topographic values except J45 and WTW in healthy eyes and J0, maximum corneal power point, and WTW in keratoconus eyes, with statistically significant differences in astigmatism (healthy), flat K (keratoconus), axis (keratoconus), J0, J45, and WTW (P < .05). Healthy eyes showed better agreement (95% limits of agreement: simulated K -0.13 to 0.40; steep K -0.30 to 0.59; flat K -0.29 to 0.51; astigmatism -0.60 to 0.64; J0 -1.15 to 1.13; J45 -1.10 to 1.20; maximum corneal power point -0.70 to 1.17; WTW -0.96 to 0.76 mm) than keratoconic eyes (simulated K -2.84 to 4.55; steep K -2.80 to 5.21; flat K -3.68 to 4.70; astigmatism -1.90 to 2.95; J0 -2.85 to 3.20; J45 -3.21 to 3.05; maximum corneal power point -7.00 to 4.51 D; WTW -1.00 to 0.88).

CONCLUSIONS

Healthy eyes showed better agreement than keratoconus eyes between Placido-based and dual-Scheimpflug topography. Both instruments could be used interchangeably with caution in healthy eyes, but not in keratoconus management.

FINANCIAL DISCLOSURE: None of the authors has a financial or proprietary interest in any material or method mentioned.

Для цитирования: Ortiz-Toquero Sara, Zuñiga Virginia, Rodriguez Guadalupe, de Juan Victoria, Martin Raul. Agreement of corneal measurements between dual rotating Scheimpflug-Placido system and Placido-based topography device in normal and keratoconus eyes.. Journal of cataract and refractive surgery. 2016. DOI: 10.1016/j.jcrs.2016.04.033.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.