Feasibility of Ultrafast Doppler technique for cranial ultrasound in neonates.
Аннотация
AIMS: The aim of this study was to compare the performances of Ultrafast Doppler ultrasound (US) with classic Doppler US, for cranial ultrasound in neonates.
MATERIALS AND
METHODS
We measured the peak systolic velocity (PSV), end-diastolic velocity (EDV) and resistive index (RI) of the anterior cerebral artery (ACA), middle cerebral artery (MCA) and posterior cerebral artery (PCA) in neonates using both conventional and Ultrafast Doppler US and acquisition times were compared. Distal ACA branches were assessed with Ultrafast Doppler US.
RESULTS
A total of 138 neonates were included. The PSV and EDV of the cranial arteries were comparable between the two Doppler methods (PSV, 64.6-85.5 cm/s vs. 63.4-84.1 cm/s, p=0.100-0.510; EDV, 19.1-26.5 cm/s vs. 17.8-24.2 cm/s, p=0.100-0.981). The RIs of the ACA and PCA were not significantly different (0.69-0.73 vs 0.68-0.74, p=0.174-0.810). Ultrafast Doppler US required shorter acquisition times than conventional Doppler US (6.7 s vs. 11.0 s, p=0.003). The PSV and EDV of the distal ACA were higher than the proximal ACA (20.1-63.3 cm/s vs. 9.4-36.7, p<0.001) although the RI was similar (0.69 vs. 0.68, p=0.251).
CONCLUSIONS
Ultrafast Doppler US provides comparable values to conventional Doppler US with shorter acquisition times. This novel imaging technique provides quantitative information and is suitable for distal cranial artery evaluation.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.