Accuracy (trueness and precision) of two handheld scanners and a smartphone for three-dimensional facial scanning: An in vivo study.

Erkizia Gaizka, Amezua Xabier, Iturrate Mikel, Solaberrieta Eneko

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The aim of this in vivo study was to evaluate and compare the accuracy (trueness and precision) of three portable facial scanning methods: a professional handheld structured-light scanner (P-HS), a low-cost handheld structured-light scanner (LC-HS), and a smartphone with a depth-sensing camera (SP).

MATERIALS AND

METHODS

Twelve adult volunteers were recruited. Seventeen facial soft-tissue landmarks were marked with adhesive fiducials, and 12 inter-landmark distances were measured. Ten repeated facial scans were acquired from each participant with each method, plus reference scans from an industrial reference scanner. The relative error (RE) was calculated, and trueness and precision were expressed as the mean (Mean-RE) and standard deviation (SD-RE) of REs per participant, distance, and method. The methods were compared using linear mixed-effects models, with pairwise comparisons adjusted by the Holm method (α = 0.05).

RESULTS

Statistically significant differences were found in the trueness and precision. The P-HS was the most accurate. The P-HS and SP showed comparable precision, about twice that of the LC-HS, which was the least precise and tended to overestimate. Despite a near-zero net trueness, the SP produced the largest deviations at individual distances. Differences between the trueness and precision of P-HS and SP were not statistically significant.

CONCLUSIONS

The facial scanning method significantly affected accuracy. The P-HS best suits high-fidelity workflows; the LC-HS remained within clinically acceptable limits in other studies for less demanding tasks; and the SP was least accurate where point-to-point fidelity is required. Method selection should follow the accuracy each workflow requires.

Для цитирования: Erkizia Gaizka, Amezua Xabier, Iturrate Mikel, Solaberrieta Eneko. Accuracy (trueness and precision) of two handheld scanners and a smartphone for three-dimensional facial scanning: An in vivo study.. Journal of prosthodontics : official journal of the American College of Prosthodontists. 2026. DOI: 10.1111/jopr.70228.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.