Sheath Rupture in Figure-Of-Eight Suturing: A Case Report and Risk Reduction Strategies.

Kobayashi Takahiro, Oginosawa Yasushi, Nakamura Yuki, Miyamoto Taro, Yamagishi Yasunobu, Hayashi Katsuhide, Kawakami Kazunobu, Kataoka Masaharu

СтатьяПеревод пока не подготовлен

Аннотация

To prevent sheath perforation during figure-of-eight suturing, operators should optimize needle placement and suture tension, avoiding proximal sheath puncture. Polyether block amide sheath usage and intraluminal device retention during suturing provide additional protection. Standardized training and prompt suture removal upon resistance during sheath withdrawal may enhance safety.

Для цитирования: Kobayashi Takahiro, Oginosawa Yasushi, Nakamura Yuki, Miyamoto Taro, Yamagishi Yasunobu, Hayashi Katsuhide, Kawakami Kazunobu, Kataoka Masaharu. Sheath Rupture in Figure-Of-Eight Suturing: A Case Report and Risk Reduction Strategies.. Clinical case reports. 2026. DOI: 10.1002/ccr3.72934.