DOI 10.1007/s00134-007-0850-5

Медицинская библиотека · Intensive care medicine · 2007

Smart Bag vs. Standard bag in the temporary substitution of the mechanical ventilation.

Lovat Robin, Watremez Christine, Van Dyck Michel, Van Caenegem Olivier, Verschuren Franck, Hantson Philippe, Jacquet Luc-Marie

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To compare in intubated patients manually ventilated in order to mirror the ventilator, the respiratory and hemodynamic effects induced by a bag device equipped with an inspiratory gas flow-limiting valve (Smart Bag, 0-Two Medical Technologies Inc., Mississauga, ON, Canada) and a Standard bag.

DESIGN

Non-randomized crossover study comparing 13 respiratory and eight hemodynamically paired parameters. Eight intubated patients were manually ventilated, each by three different intensive care workers yielding 24 sets of data for comparison. Data were collected during two sessions of manual ventilation, first with the Standard bag and then with the Smart Bag. Between each session, the patient was reconnected to the ventilator until return to the baseline. Patients, included after coronary surgery, were sedated and paralyzed.

SETTING

Intensive Care Unit, university hospital.

RESULTS

Compared with Standard bag, the Smart Bag provided a decrease of inspiratory flow (23 +/- 4.7 vs. 47.3 +/- 16.5 l/min) with a decrease of peak pressure (13.3 +/- 2.9 vs. 21.9 +/- 7.3 cmH2O) and tidal volume (9.4 +/- 2.8 vs. 12.4 +/- 2.7 ml/kg). While the expiratory time was similar, the inspiratory time increased (1.83 +/- 0.58 vs. 1.28 +/- 0.46 s) with the Smart Bag, limiting the respiratory rate (14 +/- 5 vs. 17 +/- 6 cycles/min) and the minute volume (8.8 +/- 2.9 vs. 14.4 +/- 4.9 l/min). Finally, it limited the fall of the ETCO2 (27.9 +/- 5.1 vs. 24.3 +/- 5.7 mmHg) and probably the risks of severe respiratory alkalosis. The bags similarly affected hemodynamic states.

CONCLUSION

In intubated patients manually ventilated, the Smart Bag limits the risks of excessive airway pressure and the fall of the ETCO2, with hemodynamic effects similar to those of the Standard bag.

Для цитирования: Lovat Robin, Watremez Christine, Van Dyck Michel, Van Caenegem Olivier, Verschuren Franck, Hantson Philippe, Jacquet Luc-Marie. Smart Bag vs. Standard bag in the temporary substitution of the mechanical ventilation.. Intensive care medicine. 2007. DOI: 10.1007/s00134-007-0850-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.