Clinica chimica acta; international journal of clinical chemistry · 2018

Is capillary zone electrophoresis a suitable method for estimating serum albumin: A comparison of four methods.

Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Padelli Maël, Leven Cyril, Le Reun Eymeric, Labouret Théo, Galakhoff Nicolas, Labarre Marion, Rouillé Amélie, Kerspern Hélène, Lapègue Marion, Chauvet Juliette, Plée-Gautier Emmanuelle, Carré Jean-Luc

Аннотация

BACKGROUND: The capillary zone electrophoresis method of albumin measurement is frequently used for oncologic and haematologic patients but few data exist about the agreement between the albumin measurements performed by capillary zone electrophoresis and other methods. The aim of this study was to analyse the agreement between human serum albumin measurements by capillary zone electrophoresis and by the nephelometry, bromocresol purple and turbidimetry methods.

METHOD: We prospectively measured 100 freshly collected non-frozen patient serum samples, by using four different methods: the capillary zone electrophoresis method performed with a CAPILLARYS 2 instrument, the bromocresol purple dye method performed with an Advia XPT analyser, the nephelometric method performed with a BN ProSpec analyser and the turbidimetric method with reagents from DiAgam and performed with the Advia XPT analyser.

RESULTS: A bias towards higher values in the lower range of albumin concentrations was observed with capillary zone electrophoresis compared to immunonephelometry: correlation coefficient r2 = 0.925; slope of 0.86 (0.82-0.89, 95% confidence interval), which is significantly different from 1; and an intercept of 4.94 g/L (3.67-6.16, 95% confidence interval). Similar results were observed when comparing capillary zone electrophoresis to the bromocresol purple and immunoturbidimetry methods. The capillary electrophoresis method overestimated low albumin levels by up to 25% (5 g/L).

CONCLUSION: Compared to the nephelometry, turbidimetry and bromocresol purple methods, the capillary zone electrophoresis method tends to overestimate human serum albumin concentrations for levels below 30 g/L. This discrepancy could lead to an overestimation of the nutritional status, an inappropriate scoring of the disease and a delay in nutritional treatment.