DOI 10.3109/08820538.2015.1046557
Diagnostic Consistency and Relation Between Optical Coherence Tomography and Standard Automated Perimetry in Primary Open-Angle Glaucoma.
Аннотация
PURPOSE: To assess diagnostic consistency and relation between spectral-domain optical coherence tomography (SD-OCT) and standard automated perimetry (SAP) in patients with primary open-angle glaucoma (POAG).
METHODS
This retrospective study comprised 51 eyes of 51 patients with a confirmed diagnosis of POAG. The qualitative and quantitative SD-OCT parameters (retinal nerve fiber layer thicknesses [RNFL; average, superior, inferior, nasal and temporal], RNFL symmetry, rim area, disc area, average and vertical cup/disc [C/D] ratio and cup volume) were compared with parameters of SAP (mean deviation, pattern standard deviation, visual field index, and glaucoma hemifield test reports).
RESULTS
Fifty-one eyes of 51 patients with POAG were recruited. Twenty-nine eyes (56.9%) had consistent RNFL and visual field (VF) damage. However, nine patients (17.6%) showed isolated RNFL damage on SD-OCT and 13 patients (25.5%) had abnormal VF test with normal RNFL. In patients with VF defect, age, average C/D ratio, vertical C/D ratio, and cup volume were significantly higher and rim area was lower when compared to those of the patients with normal VF. In addition to these parameters, worsening in average, superior, inferior, and temporal RNFL thicknesses and RNFL symmetry was significantly associated with consistent SD-OCT and SAP outcomes.
CONCLUSIONS
In routine practice, patients with POAG can be manifested with inconsistent reports between SD-OCT and SAP. An older age, higher C/D ratio, larger cup volume, and lower rim area on SD-OCT appears to be associated with detectable VF damage. Moreover, additional worsening in RNFL parameters might reinforce diagnostic consistency between SD-OCT and SAP.