Correlation between Central Corneal Thickness and Intraocular Pressure Measured by Goldmann Applanation Tonometry or Pascal Dynamic Contour Tonometry.
Аннотация
PURPOSE: The aim of this study was to investigate the relationship between central corneal thickness and intraocular pressure measured by Goldmann applanation tonometry and Pascal dynamic contour tonometry.
PATIENTS AND
METHODS
The study included 45 persons (90 eyes), divided into 4 groups: a) 10 normal volunteers (20 eyes); b) 16 patients (32 eyes) with primary open-angle glaucoma; c) 8 patients (16 eyes) with normal-tension glaucoma; and d) 11 patients (22 eyes) with ocular hypertension. Intraocular pressure was measured by Goldmann applanation tonometry and Pascal dynamic contour tonometry, and central corneal thickness was measured by ultrasound pachymetry. The relationship between intraocular pressure and central corneal thickness was evaluated.
RESULTS
Intraocular pressure was correlated positively but not strongly enough with central corneal thickness when it was measured by Goldmann applanation tonometry. On the contrary, there was no correlation between intraocular pressure and central corneal thickness when intraocular pressure was measured by Pascal dynamic contour tonometry.
CONCLUSION
Central corneal thickness is an important variable in the evaluation of intraocular pressure by Goldmann applanation tonometry. This factor does not interfere with the intraocular pressure measurements taken by Pascal dynamic contour tonometry.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.