Intraocular Pressure Before and After Corneal Refractive Surgery: A Prospective Comparison of Corvis ST and Ocular Response Analyzer.
Аннотация
PRÉCIS: The study showed that Corvis ST's biomechanical intraocular pressure (bIOP) and ocular response analyzer's (ORA) cornea-compensated intraocular pressure (IOPcc) do not agree well, before or after photorefractive keratectomy (PRK), and may not be used interchangeably. bIOP remained unchanged after PRK.
OBJECTIVE
To evaluate the agreement between the biomechanically corrected intraocular pressure (bIOP) measured by the Corvis ST and the IOPcc measured by the ORA before and after PRK.
PATIENTS AND
METHODS
In this prospective interventional study, a total of 53 patients (53 eyes) were included. Measurements were acquired using both the Corvis ST and ORA devices before and 3 months post-PRK. The agreement between the 2 devices was evaluated using limits of agreement (LoA) and Bland-Altman plots.
RESULTS
The participants had a mean age of 29.6 ± 5.21 years (range: 21 to 40), with 41 (77.4%) of them being females. After the surgery, the average change in intraocular pressure (IOP) was 0.3 ± 1.7 mm Hg for bIOP and -1.6 ± 4.0 mm Hg for IOPcc. The corresponding 95% LoA were -3.5 to 4.2 mm Hg and -9.5 to 6.3 mm Hg, respectively. The 95% LoA between bIOP and IOPcc after PRK was -2.3 to 8.5 mm Hg. Notably, the bIOP values were higher for IOPs <20 mm Hg and lower for IOPs >20 mm Hg compared with IOPcc.
CONCLUSIONS
The findings indicate a weak agreement between the Corvis ST-bIOP and the ORA-IOPcc both before and after PRK. These devices may not be used interchangeably for IOP measurement. bIOP exhibited less variation compared with the IOPcc, suggesting that the bIOP may be a better option for IOP reading after PRK.