Design and Testing of a Single-Element Ultrasound Viscoelastography System for Point-of-Care Edema Quantification.

СтатьяПоддержка правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Pitre John J, Koziol Leo B, Kruger Grant H, Vollmer Alan, Ophir Jonathan, Ammann Jean-Jacques, Weitzel William F, Bull Joseph L

Аннотация

Management of fluid overload in patients with end-stage renal disease represents a unique challenge to clinical practice because of the lack of accurate and objective measurement methods. Currently, peripheral edema is subjectively assessed by palpation of the patient's extremities, ostensibly a qualitative indication of tissue viscoelastic properties. New robust quantitative estimates of tissue fluid content would allow clinicians to better guide treatment, minimizing reactive treatment decision making. Ultrasound viscoelastography (UVE) can be used to estimate strain in viscoelastic tissue, deriving material properties that can help guide treatment. We are developing and testing a simple, low-cost UVE system using a single-element imaging transducer that is simpler and less computationally demanding than array-based systems. This benchtop validation study tested the feasibility of using the UVE system by measuring the mechanical properties of a tissue-mimicking material under large strains. We generated depth-dependent creep curves and viscoelastic parameter maps of time constants and elastic moduli for the Kelvin model of viscoelasticity. During testing, the UVE system performed well, with mean UVE-measured strain matching standard mechanical testing with maximum absolute errors ≤4%. Motion tracking revealed high correlation and signal-to-noise ratios, indicating that the system is reliable.

Предметные рубрики MeSH