Feasibility of High-Flow Oxygen Therapy in Apnea Testing for Brain Death Diagnosis.

Barrier Damien, Vanacker Ludivine, Muller Grégoire, Szychowiak Piotr, Bretagnol Anne, Levebvre Isabelle, Narcisse Éric, Pascot Léa, Boulain Thierry, Nay Mai-Anh

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study aimed to assess the feasibility and safety of use of high-flow oxygen (HFO) through the endotracheal tube for apnea testing during brain death evaluation.

DESIGN

Prospective, interventional, single-center study.

SETTING

ICUs at the Orléans University Hospital, France.

PATIENTS: All adult patients meeting the criteria for clinical brain death and requiring an apnea test were eligible.

INTERVENTIONS: Fifteen patients underwent two successive apnea tests in the same order, each lasting 10 minutes: The first apnea test used 8 L/min of oxygen via a T-piece without a positive end-expiratory pressure valve connected to the endotracheal tube (standard test). The second apnea test used HFO at 50 L/min and F io 2 of 1.0 through a specialized device connected to the endotracheal tube. Arterial blood gases were measured every 2 minutes during both apnea tests. The primary objective was to investigate the feasibility and safety of use of HFO through the endotracheal tube for apnea testing compared with use of a T-piece in patients whose clinical examination was consistent with brain death and required an apnea test.

MEASUREMENTS AND MAIN

RESULTS

All patients reached Pa co 2 greater than 60 mm Hg by the end of both apnea tests. At 10 minutes, Pa co 2 was 69.1 ± 5.9 mm Hg with the T-piece and 71.3 ± 6.0 mm Hg with HFO. Pa o 2 decreased significantly during the T-piece apnea test (from 404 ± 115 to 215 ± 118 mm Hg) but remained stable with HFO apnea test (from 342 ± 114 to 308 ± 92 mm Hg).

CONCLUSIONS

HFO through the endotracheal tube is feasible and safe method for apnea testing. In patients with confirmed brain death via the standard apnea test, HFO yielded consistent diagnostic results while maintaining better oxygenation. Further research, particularly randomized trials, is needed to confirm the benefits of HFO in apnea testing.

Для цитирования: Barrier Damien, Vanacker Ludivine, Muller Grégoire, Szychowiak Piotr, Bretagnol Anne, Levebvre Isabelle, Narcisse Éric, Pascot Léa, Boulain Thierry, Nay Mai-Anh. Feasibility of High-Flow Oxygen Therapy in Apnea Testing for Brain Death Diagnosis.. Critical care medicine. 2025. DOI: 10.1097/ccm.0000000000006717.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.