DOI 10.1186/s12886-017-0550-z

Медицинская библиотека · BMC ophthalmology · 2017

Comparison of astigmatism prediction error taken with the Pentacam measurements, Baylor nomogram, and Barrett formula for toric intraocular lens implantation.

Park Dae-Young, Lim Dong Hui, Hwang Sungsoon, Hyun Joo, Chung Tae-Young

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

To evaluate and compare the astigmatism prediction errors taken with the Pentacam measurements, Baylor nomogram, and Barrett formula for toric intraocular lens (IOL) implantation.

METHODS

Phacoemulsification with toric Precizon IOL implantation was performed in 41 eyes with corneal astigmatism (range, 1 to 5 diopters (D)) determined by IOLMaster and SimK on Pentacam. Preoperative corneal astigmatism measurements were obtained from IOLMaster readings (IOLMaster, Baylor-IOLMaster, and Barrett-IOLMaster) and Pentacam readings (SimK, Baylor-SimK, Barrett-SimK, wavefront, true net power, total corneal refractive power, and vector derived by manual vector summation using corneal front and back astigmatism). Prediction error and intraclass correlation coefficient (ICC) between the measured (or calculated) astigmatism by IOLMaster and Pentacam and the estimated corneal astigmatism estimated by IOL toricity power and residual astigmatism were determined.

RESULTS

The centroid errors in prediction error with IOLMaster, SimK, Baylor-IOLMaster, Baylor-SimK, Barrett-IOLMaster, Barrett-SimK, wavefront, true net power, total corneal refractive power, and vector were 0.59@103, 0.61 @103, 0.37@161, 0.41@162, 0.24@171, 0.36@162, 0.42@106, 0.04@8, 0.07@82, and 0.03@82, respectively, in with-the-rule (WTR) astigmatism eyes at postoperative 3-month. They were 0.22@87, 0.20@74, 0.16@21, 0.54@10, 0.43@3, 0.33@19, 0.51@25, 0.31@58, 0.29@50, and 0.14@50 in against-the-rule (ATR) astigmatism eyes. Of the ten modalities, vector showed the lowest WTR astigmatism prediction error and the highest ICC between the predicted and the estimated corneal astigmatism for both WTR and ATR eyes.

CONCLUSION

Vector summation using anterior and posterior corneal surface power taken with the Pentacam yields the least astigmatism prediction error and is a promising tool for determining toric IOL cylinder power.

Для цитирования: Park Dae-Young, Lim Dong Hui, Hwang Sungsoon, Hyun Joo, Chung Tae-Young. Comparison of astigmatism prediction error taken with the Pentacam measurements, Baylor nomogram, and Barrett formula for toric intraocular lens implantation.. BMC ophthalmology. 2017. DOI: 10.1186/s12886-017-0550-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.