DOI 10.1016/j.phro.2026.100942
An efficient automated approach for accumulated dose estimation in prostate cancer radiotherapy.
Аннотация
BACKGROUND AND PURPOSE: Anatomical variations during prostate radiotherapy can cause discrepancies between planned and delivered doses, which may become more pronounced in (ultra-)hypofractionated regimens. However, a standardized method for accumulated dose assessment is lacking. This study presented and validated an efficient automated workflow for retrospective dose accumulation in moderately hypofractionated prostate treatments, avoiding deformable image registration and daily dose recalculation.
MATERIALS AND
METHODS
Twenty prostate cancer patients were analyzed retrospectively. The workflow combined daily cone-beam computed tomography (CBCT) imaging, online rigid registration, and artificial intelligence-based auto-contouring with planned dose data to estimate accumulated doses. Dose-volume metrics were analyzed for prostate, rectum, and bladder. Anatomical variations were assessed through volume and position analysis. Method accuracy was validated against CBCT-based dose recalculations in selected cases representing minimal and maximal variations.
RESULTS
The workflow required about one minute per fraction. Interfractional volume variability (coefficient of variation) was 35.2 %, 20.6 %, and 9.5 % for bladder, rectum, and prostate, respectively. Mean prostate displacement was 3.49 ± 2.07 mm, and the dice similarity coefficient averaged 0.83 ± 0.08. Bladder dose metrics (V20Gy, V50Gy, mean) differed by > 2 % in 45-49 % of sessions, rectal metrics in 35-39 %. Accumulated dose estimates agreed within ± 2 % of CBCT-based recalculations.
CONCLUSIONS
Our automated workflow provides a robust, time-efficient method for monitoring interfractional dose variations in prostate radiotherapy. While constrained by CBCT image quality, the accuracy and practicality of the proposed method support its integration into clinical workflows, potentially improving offline adaptive decision-making.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.