DOI 10.1016/s0002-9394(14)72727-8
Calculation of intraocular lens power after radial keratotomy with computerized videokeratography.
Аннотация
PURPOSE: Because standard methods to determine intraocular lens power are not adequate in eyes that have had radial keratotomy, we undertook this study to evaluate the corneal power derived from computerized videokeratography for use in intraocular lens power calculations.
METHODS
We examined four eyes of three patients who had radial keratotomy and who underwent phacoemulsification cataract surgery with implantation of a posterior chamber intraocular lens. We used a computerized videokeratography-derived corneal curvature value in the Holladay formula for intraocular lens calculations. We determined the ideal intraocular lens power and the keratometric value that would have led to the ideal intraocular lens power from the postoperative refraction at 6.1 +/- 1.1 months after cataract extraction. The ideal keratometric value was compared with the keratometric values derived from computerized videokeratography, standard keratometry, contact lens overrefraction, and refractions before and after radial keratotomy.
RESULTS
The postoperative refraction at approximately six months averaged -0.32 +/- 0.63 diopter (range, -0.88 to +0.75 diopter) different than the aim. The mean power in ring 3, which was the closest keratometric value to the ideal, disclosed only 0.09 +/- 0.73 diopter and -0.10 +/- 0.72 diopter of deviation from the ideal keratometric and intraocular lens powers, respectively. One to two weeks after phacoemulsification cataract surgery with implantation of a posterior chamber intraocular lens, the videokeratographic differential map disclosed steepening at the wound site with variable regression by six months in all patients.
CONCLUSION
Results suggest that, after radial keratotomy, using the keratometric value derived from computerized videokeratography in intraocular lens calculations is more accurate than using keratometric values measured by routine methods.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.