DOI 10.7717/peerj.21170

Медицинская библиотека · PeerJ · 2026

Impact of water on the fracture behavior of an advanced lithium disilicate.

Lu Yuqing, Tribst João Paulo Mendes, Kodolányi János, Kleverlaan Cornelis J, Feilzer Albert J, Dal Piva Amanda Maria de Oliveira

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To investigate the influence of water on the monotonic and fatigue strength of a lithium disilicate glass-ceramic.

METHODS

Ninety bar-shaped specimens (1.0 × 1.0 × 12.0 mm) were prepared from Advanced Lithium Disilicate (CEREC Tessera; Dentsply Sirona). Half of the specimens were stored in deionized water (W) at 37 °C for 30 days, while the other specimens stayed dry (D). A three-point bending test was carried out in a dry/wet environment (subgroup d/w) to determine monotonic strength. A stepwise fatigue test was conducted for dry-stored specimens in a dry environment (Dd) and wet-stored specimens in water (Ww).

RESULTS

For monotonic strength, the testing environment had a significant effect (p < 0.001), whereas the storage environment did not (p = 0.054). Testing in water generated a lower monotonic strength (Dw: 243 ± 35 MPa; Ww: 250 ± 57 MPa) than in a dry environment (Dd: 324 ± 74 MPa; Wd: 376 ± 60 MPa). Wet storage combined with testing in water exhibited similar fatigue strength (Ww: 148 ± 39 MPa) to the group without water intervention (Dd: 152 ± 36 MPa).

CONCLUSIONS

While storage in 37 °C water for 30 days causes ion release from lithium disilicate, it does not decrease its monotonic strength. The compressive loading in water resulted in a degradation of around 30%. The fatigue protocol in this study resulted in about 50% of the initial strength regardless of testing conditions.

Для цитирования: Lu Yuqing, Tribst João Paulo Mendes, Kodolányi János, Kleverlaan Cornelis J, Feilzer Albert J, Dal Piva Amanda Maria de Oliveira. Impact of water on the fracture behavior of an advanced lithium disilicate.. PeerJ. 2026. DOI: 10.7717/peerj.21170.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.