INTRODUCTION

The prerequisite for transoral microsurgery of laryngeal lesions is an uninterrupted line of sight to the operative field. Patients with cervical stiffness or anatomical variations that prevent adequate laryngeal exposure are typically unsuitable for this procedure. In such cases, a curved video laryngoscope may facilitate improved access to the larynx. However, conventional suspension laryngoscopy setups are incompatible with standard video laryngoscopes.

OBJECTIVE

This study evaluates three attachment methods for integrating a curved video laryngoscope into a surgical setup, focusing on their resistance to external forces.

METHODS

This study assessed three different attachment methods (3D-printed clamp vs. metal bracket vs. articulated stand) for a video laryngoscope in a surgical setup. External forces, both lateral and rotational (torque), were applied and continuously measured until laryngeal visualization was compromised by displacement of the video laryngoscope.

RESULTS

The metal bracket demonstrated significantly (p < 0.001) higher resistance to lateral forces (median 184.49 N, 95% CI [181.59-189.61 N]) compared to the articulated stand (median 88.16 N, 95% CI [76.73-88.98 N]) and the 3D-printed clamp (median 55.59 N, 95% CI [54.74-57.58 N]). The articulated stand exhibited significantly (p < 0.005) greater torque resistance (median 9.57 N m, 95% CI [5.65-9.87 N m]) compared to the metal bracket (median 1.58 N m, 95% CI [1.57-2.13 N m]) and the 3D-printed clamp (median 2.46 N m, 95% CI [2.24-2.79 N m]).

CONCLUSION

Overall, the articulated stand outperformed the other attachment methods, displaying robust resistance to lateral forces and superior rotational stability.

LEVEL OF EVIDENCE: Level 4.

Для цитирования: Kienle Linus L, Schild Leon R, Seitz Andreas M, Hahn Viola D, Greve Jens, Hoffmann Thomas K, Schuler Patrick J, Boehm Felix. A Comparative Evaluation of Three Different Attachment Methods for a Video Laryngoscope in a Surgical Setting.. Laryngoscope investigative otolaryngology. 2025. DOI: 10.1002/lio2.70172.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.