DOI 10.1016/j.ogla.2024.09.005
Manometric Intraocular Pressure Reduction with Negative Pressure Using Ocular Pressure Adjusting Pump Goggles.
Аннотация
OBJECTIVE
The ocular pressure adjusting pump is a novel noninvasive and nonpharmacological intraocular pressure (IOP)-lowering device consisting of goggles attached to a negative pressure (NP) pump. This study aimed to determine the effect of NP in ocular pressure adjusting pump goggles on IOP using continuous direct manometry.
DESIGN
Prospective, single-arm, single center, basic physiological research trial.
SUBJECTS: Adult patients undergoing cataract surgery.
METHODS
Direct manometry was performed via an anterior chamber cannula attached to an IOP sensor. This was placed just prior to each subject's previously scheduled cataract surgery. Intraocular pressure was continuously monitored every 0.5 seconds through the following sequence of 30 seconds each: baseline IOP measurement, NP -10 mmHg, NP off, NP -20 mmHg, and NP off.
MAIN OUTCOME MEASURES: Intraocular pressure response to application of NP.
RESULTS
All 17 subjects had a dose-dependent reduction in IOP, with a mean IOP decrease from 16.9 to 11.3 (33%) with -10 mmHg of NP and from 15.7 to 7.7 (51%) mmHg with -20 mmHg of NP. Intraocular pressure returned to baseline for each cycle when NP was discontinued. There were no patients with increased IOP.
CONCLUSIONS
Negative pressure with the ocular pressure adjusting pump results in a consistent dose-dependent reduction in IOP, with return to baseline IOP when NP is discontinued.
FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.