DOI 10.1111/jce.70049

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2025

Hemolysis Induced by Pulsed-Field Ablation of Atrial Arrhythmias: A Comparative Analysis of Current Systems.

Bruss Johannes, Kueffer Thomas, Tanner Hildegard, Noti Fabian, Haeberlin Andreas, Thalmann Gregor, Kozhuharov Nikola Asenov, Kovacs Boldizsar, Spahiu Valon, Siklody Claudia Herrera, Reichlin Tobias, Roten Laurent

СтатьяСравнительное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

INTRODUCTION

Pulsed-field ablation (PFA) is an emerging technology associated with dose-dependent hemolysis as a recently recognized side effect. This study aimed to compare hemolysis levels and assess dose-dependency across three PFA systems: a pentaspline catheter (PSC), a lattice-tip focal catheter (LTFC), and a variable loop circular catheter (VLCC).

METHODS

Patients treated for atrial arrhythmias with the LTFC (n = 29) or the VLCC (n = 30) were included from a prospective registry. A matched cohort of patients treated with the PSC (n = 28) was recruited from the same registry. Creatinine levels and markers of hemolysis were measured pre-ablation and 1 day postablation.

RESULTS

Haptoglobin levels decreased significantly more with the PSC and VLCC compared to the LTFC (-0.65 [-0.76, -0.49] g/L; -0.56 [-0.78, -0.43] g/L, -0.21 [-0.32, -0.1] g/L, respectively; p < 0.001 for both). Per-application decreases in haptoglobin also differed (-17.5 [-20.38, -13.58] mg/L, -24.35 [-36.36, -17.92] mg/L, -3.61 [-5.98, -2.13] mg/L, respectively; p < 0.001 for both). There was no significant difference in haptoglobin decrease between the PSC and VLCC per procedure (p = 1.0). Haptoglobin decrease per application was significantly larger with the VLCC compared to the PSC (p = 0.0048). Per procedure LDH increase followed a similar trend (49 [18, 81.25] U/L; 14 [6, 60] U/L; 13 [-4, 46] U/L; respectively; p = 0.037). No hemolysis-related complications were observed.

CONCLUSIONS

Hemolysis levels vary significantly among PFA platforms. Focal PFA catheters induce less hemolysis per procedure and application compared to large-footprint catheters.

Для цитирования: Bruss Johannes, Kueffer Thomas, Tanner Hildegard, Noti Fabian, Haeberlin Andreas, Thalmann Gregor, Kozhuharov Nikola Asenov, Kovacs Boldizsar, Spahiu Valon, Siklody Claudia Herrera, Reichlin Tobias, Roten Laurent. Hemolysis Induced by Pulsed-Field Ablation of Atrial Arrhythmias: A Comparative Analysis of Current Systems.. Journal of cardiovascular electrophysiology. 2025. DOI: 10.1111/jce.70049.