DOI 10.3174/ajnr.a9587

Медицинская библиотека · AJNR. American journal of neuroradiology · 2026

An Unusual Stroke Mimicker: A Case of CT Artifact Produced by Contrast Leakage onto the Detector Panel.

Hwang Anna J, Zinger Amanda, Khan Nazir

СтатьяПеревод пока не подготовлен

Аннотация

To improve image quality and reduce the possibility of misdiagnosis, radiologists should be informed of both common and uncommon manifestations of artifact on computed tomography (CT) imaging. We present an unusual case of scanner-based artifact, which produced peripheral dark attenuation defects on several cranial CT exams obtained from a single scanner, mimicking cerebral infarction. After confirmation of the artifact with the use of a water-based phantom, the source was found to be from contrast material which had leaked onto the CT detector panel. The ability to recognize artifactual patterns and an understanding of quality control measures are crucial to identify and mitigate sources of artifact.

Для цитирования: Hwang Anna J, Zinger Amanda, Khan Nazir. An Unusual Stroke Mimicker: A Case of CT Artifact Produced by Contrast Leakage onto the Detector Panel.. AJNR. American journal of neuroradiology. 2026. DOI: 10.3174/ajnr.a9587.